ORIGINAL ARTICLE |
|
Year : 2016 | Volume
: 10
| Issue : 2 | Page : 162-166 |
|
Diaphragmatic ultrasound as a predictor of successful extubation from mechanical ventilation: thickness, displacement, or both?
Ayman I Baess MD, PhD 1, Tamer H Abdallah2, Doaa M Emara3, Maged Hassan1
1 Department of Chest Diseases, Alexandria Faculty of Medicine, Alexandria, Egypt 2 Department of Critical Care Medicine, Alexandria Faculty of Medicine, Alexandria, Egypt 3 Department of Diagnostic & Interventional Radiology, Alexandria Faculty of Medicine, Alexandria, Egypt
Date of Submission | 19-Nov-2015 |
Date of Acceptance | 28-Dec-2015 |
Date of Web Publication | 23-Jun-2016 |
Correspondence Address: Ayman I Baess Department of Chest, Secretary Office, Alexandria Faculty of Medicine, Alexandria 21613 Egypt
Source of Support: None, Conflict of Interest: None | Check |
DOI: 10.4103/1687-8426.184370
Background: Best predictor of successful extubation after mechanical ventilation is a matter of debate. Objective: The aim of this study was to assess whether the degree of diaphragm thickening and/or diaphragm displacement (DD) as measured by means of ultrasound during a weaning trial can predict extubation outcomes. Methods: Thirty patients who were planned for weaning after being intubated and mechanically ventilated were prospectively enrolled in the study between January and June 2015. The rapid-shallow breathing index (RSBI) was subsequently calculated, and diaphragmatic ultrasound was then carried out to assess DD and diaphragm thickening during tidal inspiration. The primary outcome was extubation success or failure. Results: Of the 30 patients included in the study, 15 were male. The mean age of patients was 59.17+13.17 years. The median duration of intubation before weaning was 4 days. There was a significant difference between mean inspiratory and expiratory diaphragmatic thickness (TDI) (t=9.66, P<0.001). An receiver operating characteristic curve was constructed for the end inspiratory TDI, end expiratory TDI, delta TDI, DD, and RSBI. The RSBI performed better than all other parameters, with an area under the curve of 0.968. A cut-off value of 73.5 had 87% sensitivity and 100% specificity for predicting extubation success. All other parameters had an area under the curve less than 0.7. (0.559, 0.624, 0.655, and 0.512 for end inspiratory TDI, end expiratory TDI, delta TDI, and diaphragmatic displacement, respectively). Conclusion: Sonographically measured TDI performed better than displacement in predicting value for weaning outcome. In a respiratory ICU, however, the RSBI seems to be a more reliable and accurate tool for the purpose and should be considered in every weaning protocol. Whether TDI can be evaluated using low-frequency ultrasound probes needs to be validated by further studies. Keywords: diaphragm ultrasound, extubation, predictor
How to cite this article: Baess AI, Abdallah TH, Emara DM, Hassan M. Diaphragmatic ultrasound as a predictor of successful extubation from mechanical ventilation: thickness, displacement, or both?. Egypt J Bronchol 2016;10:162-6 |
How to cite this URL: Baess AI, Abdallah TH, Emara DM, Hassan M. Diaphragmatic ultrasound as a predictor of successful extubation from mechanical ventilation: thickness, displacement, or both?. Egypt J Bronchol [serial online] 2016 [cited 2024 Mar 28];10:162-6. Available from: http://www.ejbronchology.eg.net/text.asp?2016/10/2/162/184370 |
Introduction | | |
Difficulty in weaning from mechanical ventilation (MV) is one of the most frequently encountered problems in modern ICUs. An estimated 20% of mechanically ventilated patients face failed extubation (requiring reintubation within 48 h of extubation) [1]. A trial of failed extubation induces several detrimental consequences, including cardiorespiratory stress, prolonged ICU stay, and increased mortality [1],[2],[3]. Delay in attempting extubation also increases the likelihood that patients would face the inherent risks of MV, including ventilator-associated pneumonia and ventilator-induced diaphragmatic atrophy [4].
Attempts have been ongoing to devise tools that can accurately determine the ideal timing of extubation. Among these tools are measuring the minute ventilation, maximal inspiratory pressure, and tracheal occlusion pressure at 0.1 s with variable degrees of success [5],[6]. Another predictive tool, the rapid-shallow breathing index (RSBI) has gained popularity as a more accurate index in predicting success of extubation, but this ability is limited in patients weaned through pressure support (PS) [7],[8].
The diaphragm plays a central role in the process of spontaneous ventilation, and it seems intuitive that, in patients receiving MV, a properly functioning diaphragm should herald successful weaning. Although the maximal inspiratory pressure is an indirect method of assessing the diaphragm, the more traditional methods of studying diaphragmatic dysfunction include fluoroscopy, phrenic nerve conduction study, and transdiaphragmatic pressure measurement [6]. These latter methods have serious limitations, including the use of ionizing radiation, invasiveness, relative unavailability, and sometimes the need for patient transportation [6],[9]. Ultrasound has emerged as a cheap, widely available, free-from-radiation, bed-side tool for assessment of the characteristics of diaphragmatic movement, such as amplitude, force, and velocity of contraction, special patterns of motion, and changes in diaphragmatic thickness (TDI) during inspiration [9],[10]. Many of these parameters have been studied as a guidance in predicting the course of weaning from MV [6],[7],[9],[10],[11],[12].
This study aimed to assess whether the degree of diaphragm thickening (DT) and/or diaphragm displacement (DD) as measured using ultrasound during a weaning trial can predict extubation outcomes.
Methods | | |
Patients
Thirty patients were prospectively enrolled in the study between January and June 2015. Patients were intubated and mechanically ventilated in the general and respiratory ICU in Alexandria Main University Hospital. The study was approved by the Ethics Committee of the Faculty of Medicine, Alexandria University. Patients who were planned for weaning either using low-level PS or spontaneous breathing trial were included if they fulfilled the following criteria: age above 18 years, hemodynamic stability without need for vasopressors, and a duration of continuous MV of 30 days or greater. Patients were excluded if the cause for intubation was upper airway obstruction or there was pregnancy, dressing on the right lower rib cage hindering sonographic access, recent intra-abdominal surgery, a history of empyema, or pleurodesis.
Study design
Eligible patients were subjected to full history taking and clinical examination after informed consent was obtained from the patient or their next of kin. The tidal volume and breathing frequency were recorded on the least PS the patient was given and the RSBI was subsequently calculated [frequency/tidal volume (l)]. Diaphragmatic ultrasound was then performed to assess the degree of excursion/displacement (DD) and muscle thickening (DT) during tidal inspiration. The procedure was performed by physicians who had no role in the management of patients. Sonographic examination was carried out for spontaneous breathing trial patients when they were disconnected from ventilator. For the PS group, diaphragm ultrasound was performed when the patients were still on ventilatory support. Weaning then ensued. The primary outcome was extubation success or failure. Patients were divided into two groups: the successful group, which included patients who could sustain spontaneous breathing for more than 48 h following extubation, and the failure group (FG), which included patients who required reintubation or noninvasive ventilation within 48 h after extubation.
Diaphragm ultrasound
The right diaphragmatic cupola was examined with a 2–4 MHz phased-array probe (Philips Healthcare, Andover, Massachusetts, USA). The right side was chosen due to the better sonographic window provided by the liver [13]. Patients were seated in a semirecumbent position and the probe was put either immediately below the right costal margin at the midclavicular line or on the last two spaces at the anterior axillary line. The probe was then directed backward, caudally, and slightly medially until the dome of the diaphragm was visualized [9]. At B-mode, the TDI at the dome was measured in mm at end-inspiration and end-expiration. Consequently, the delta TDI was calculated as percentage from the following formula:
Subsequently, using M-mode, the excursion of the diaphragm (DD) at the dome was also measured in centimeters. For each of the measurements, five readings were taken and the average was recorded.
Statistical analysis
Data were fed to the computer and analyzed using IBM SPSS software package, version 20.0. Qualitative data were described using number and percentage. Normally distributed quantitative data were described using mean±SD, whereas non-normally distributed data were expressed in range (minimum–maximum) and median. The Mann–Whitney test (Z) was used for non-normally distributed quantitative variables to compare the success and failed groups, whereas Pearson’s coefficient (r) was calculated to correlate between two normally distributed quantitative variables. Significance of the obtained results was judged at the 5% level.
Results | | |
Of the 30 patients included in the study, 15 were male. The mean age of patients was 59.17+13.17 years. [Table 1] and [Figure 1] demonstrate the underlying causes that lead to intubation in the studied patients. Fifteen patients (50%) suffered from chronic airway obstruction in the form of chronic obstructive pulmonary disease or bronchiectasis. Only 13% of patients had an extrathoracic problem (opiate poisoning, cerebrovascular stroke, and polytrauma). The median duration of intubation before weaning was 4 days (1–30 days). Seventeen patients were weaned using PS and 13 using T-piece. The sonographic and clinical measurements for patients are shown in [Table 1]. Successful weaning was achieved in 23 patients (76.7%). The mean duration of MV in the successful group was 4 days (1–30 days), whereas it was 7 days (3–18 days) in the FG. The difference was not found to be statistically significant (P=0.06). There was a significant difference between mean inspiratory and expiratory TDI (t=9.66, P<0.001). An receiver operating characteristic curve was constructed for the end inspiratory TDI, end expiratory TDI, delta TDI, DD, and RSBI [Figure 2]. The areas under the curve (AUC), sensitivity, specificity, positive predictive value, negative predictive value, and proposed cutoffs are reported in [Table 2]. The RSBI performed better than all other parameters, with an AUC of 0.968. A cut-off value of 73.5 had 87% sensitivity and 100% specificity for predicting extubation success. All other parameters had an AUC less than 0.7. | Table 1 Distribution of the studied cases according to different studied parameters (N=30)
Click here to view |
| Figure 1: Underlying disease in mechanically ventilated patients in both groups. COPD, chronic obstructive pulmonary disease; OHS, obesity hypoventilation syndrome.
Click here to view |
| Figure 2: Receiver operating characteristic curve for different parameters to predict outcome (success). RSBI, rapid-shallow breathing index.
Click here to view |
| Table 2 Agreement (sensitivity, specificity, and accuracy) for different parameters to predict outcome (success)
Click here to view |
Discussion | | |
Since its development by Yang and Tobin [14], the RSBI has gained considerable popularity as a reliable tool for predicting weaning from MV. Given the central role of the diaphragm in the process of ventilation and the widespread availability of ultrasound machines in the ICUs, attempts have been made to gauge the utility of diaphragmatic ultrasound in guiding the weaning procedure. Diaphragmatic displacement [7],[10],[12] and more recently diaphragmatic thickening [6],[11] as measured using ultrasound are now used as a surrogate for diaphragmatic function. At our institution, the RSBI is a major determinant of the weaning decision. The current study was designed to assess whether DT and DD can be reliable like RSBI in terms of ability to predict the outcome of weaning. Both DT and DD were assessed using a low-frequency ultrasound probe.
Diaphragmatic excursion is the final product of diaphragmatic strength. As poor endurance is an important cause of failed weaning, evaluation of diaphragm movement using ultrasonography arises as an important tool. In this study, a low-frequency probe was used to assess the dome of the right hemidiaphragm at M-mode. Both B and M modes can be used for this purpose [9],[13]. In a large study conducted on 210 healthy individuals, Boussuges et al. [13] used M-mode ultrasonography to determine normal values for diaphragmatic excursion. It was found that the lower limit for DD during quite and deep breathing were 1 and 4.7 cm, respectively. In the present study, the mean DD was l.5 cm [Table 1], which slightly exceeds values for normal individuals. Half of the studied patients had chronic airway obstruction. It has been shown that, when corrected for lung volume, the contractile strength of the diaphragm in chronic obstructive pulmonary disease is not reduced compared with controls [15] and may even be enhanced in some cases [16],[17]. Lerolle et al. [10] evaluated the role of ultrasound in detecting diaphragmatic dysfunction in intubated postcardiac surgery patients and found that DD less than 2.5 cm at maximal breathing effort was a predictor of prolonged intubation. Kim et al. [12] determined that a DD less than 1 cm at tidal breathing is indicative of diaphragmatic dysfunction and predictive of weaning failure. For the specific goal of determining a cut-off point to predict successful weaning, Jiang et al. [7] studied 55 intubated patients and suggested 1.1 cm or more at quiet breathing as a cut-off value for successful weaning. Proponents of this index for prediction of weaning argue that it is more accurate compared with the RSBI [7],[11]. Their explanation is that patients who sustain tidal volumes using accessory respiratory muscles before extubation, fatigue and fail extubation, which is why, they argue, relying on RSBI can be misleading [7],[11]. As much as this holds true, the opposite is possible. Although diaphragmatic strength is only one factor for successful extubation (as measured by DD), increased respiratory load might lead to failure, compromising the predictive strength of displacement (RSBI expresses the end product of the balance between strength and load). The present study did not find DD a useful parameter in predicting weaning outcome (AUC 0.51, P=0.92) when compared with the RSBI (AUC 0.97, P<0.001). In total, 87% of the studied patients suffered from a thoracic problem (imposing large ventilatory load), which might be why the RSBI performed better than DD. Moreover, the small size of the FG (seven of 30) may have obscured the relation between DD and weaning outcome.
Since its first description [18], sonographic DT evaluation has been carried out using high-frequency probes at the zone of apposition [9]. B-mode has been preferred over M-mode for evaluating DT [9]. Two studies (with different methodologies) have recently evaluated the concept of using DT (a surrogate for diaphragmatic strength) as a predictor index for weaning outcome. DiNino et al. [11] examined 63 intubated patients with ultrasonography before their first weaning trial and determined the degree of DT during tidal breathing. They found that a cutoff of 30% or more for DT was a good predictor for weaning success. However, Ferrari et al. [6], in their study on 46 tracheostomized patients who failed weaning at least once and underwent sonographic evaluation of DT during maximal inspiration during a weaning trial, found that a cut-off value of 36% was associated with successful weaning. Despite the different cut-off values, the two groups agreed that DT is more accurate compared with the RSBI in predicting successful extubation. They held a similar argument that the RSBI is a product of the work of diaphragm and accessory muscles (which are fatigable), whereas DT expresses the degree of the more sustainable power.
In our study, we tried to use the low-frequency probe − the available probe in our institution − to determine DT at frozen B-mode images during tidal breathing. As a proof of concept, DT could be evidenced using this method. Our results show a significant difference between inspiratory and expiratory TDI (means 0.67 and 0.52 cm, respectively, P<0.001). These values, however, are larger than the values reported from previous studies (0.24 to 0.35 cm at extremes of lung volumes) [6]. This is probably caused by the much lower resolution of low-frequency probes that make determining the exact boundaries of the muscle sheet challenging. Our data have shown that the values of DT had an AUC of 0.65, which means it can be predictive of weaning outcome with fair accuracy. Using a cut-off value of 30% or more, outcome could be predicted with 69% sensitivity and 71% specificity. This cutoff was identical to the one reported by DiNino et al. [11] but was smaller than that reported by Ferrari et al. [6] (36%). As previously discussed, this discrepancy is due to the different respiratory maneuver followed during the measurement. The rather lower sensitivity and specificity of DT in our studied patients might be due to the small size of the FG, which may have blurred the differences between the groups. Moreover, the disease profile of the studied cohort might have had an effect. It is noteworthy, however, that the range of DT found in the present study (0.16–0.62) was not far away from the values reported from researchers who used high-frequency ultrasound probe to detect DT (0.42–0.74) [18]. We thus can only conclude that examining DT using low-frequency probes needs further probing.
Among the challenges of using ultrasound-derived measurements to decide on the status of ventilatory apparatus is the considerable variability of results between different operators. It is a well-known caveat that results of ultrasonography are operator-dependent. We cannot exclude this element from being an obstacle toward getting a reproducible cut-off point to predict a certain outcome in the ICU. Another challenge inherent to measurements of minute structure (in fraction of cm) is the difficulty of reproducing exact results for serial measurements even by the same operator. With that said, our experience shows that, with enough practice, both interoperator and intraoperator fallacies can be fairly overcome.
This study has limitations. Sonographic measurements of the diaphragm were not complemented with more direct measurements of function (such as the maximal inspiratory pressure and transdiaphragmatic pressure). The small size of the FG and the relative homogeneity of the studied cohort do not allow generalization of the results.
Conclusion | | |
Sonographically measured diaphragmatic thickening performed better than displacement in predicting value for weaning outcome. In a respiratory ICU, however, the RSBI seems to be a more reliable and accurate tool for the purpose and should be considered in every weaning protocol. Whether DT can be evaluated using low-frequency ultrasound probes needs to be validated by further studies.
Acknowledgements | | |
Conflicts of interest
There are no conflict of interest.
References | | |
1. | Epstein SK. Extubation. Respir Care 2002; 47(4):483–492. discussion 493–495. |
2. | Esteban A, Anzueto A, Frutos F, Alía I, Brochard L, Stewart TE et al. Characteristics and outcomes in adult patients receiving mechanical ventilation: a 28-day international study. JAMA 2002; 287(3):345–355. |
3. | Funk GC, Anders S, Breyer MK, Burghuber OC, Edelmann G, Heindl W et al. Incidence and outcome of weaning from mechanical ventilation according to new categories. Eur Respir J 2010; 35(1):88–94. |
4. | Hudson MB, Smuder AJ, Nelson WB, Bruells CS, Levine S, Powers SK. Both high level pressure support ventilation and controlled mechanical ventilation induce diaphragm dysfunction and atrophy. Crit Care Med 2012; 40(4):1254–1260. |
5. | El-Khatib MF, Bou-Khalil P. Clinical review: liberation from mechanical ventilation. Crit Care 2008; 12(4):221. |
6. | Ferrari G, De Filippi G, Elia F, Panero F, Volpicelli G, Aprà F. Diaphragm ultrasound as a new index of discontinuation from mechanical ventilation. Crit Ultrasound J 2014; 6(1):8. |
7. | Jiang JR, Tsai TH, Jerng JS, Yu CJ, Wu HD, Yang PC. Ultrasonographic evaluation of liver/spleen movements and extubation outcome. Chest 2004; 126(1):179–185. |
8. | Lee KH, Hui KP, Chan TB, Tan WC, Lim TK. Rapid shallow breathing (frequency-tidal volume ratio) did not predict extubation outcome. Chest 1994; 105:540–543. |
9. | Matamis D, Soilemezi E, Tsagourias M, Akoumianaki E, Dimassi S, Boroli F et al. Sonographic evaluation of the diaphragm in critically ill patients. Technique and clinical applications. Intensive Care Med 2013; 39(5):801–810. |
10. | Lerolle N, Guérot E, Dimassi S, Zegdi R, Faisy C, Fagon JY, Diehl JL. Ultrasonographic diagnostic criterion for severe diaphragmatic dysfunction after cardiac surgery. Chest 2009; 135(2):401–407. |
11. | DiNino E, Gartman EJ, Sethi JM, McCool FD. Diaphragm ultrasound as a predictor of successful extubation from mechanical ventilation. Thorax 2014; 69(5):423–427. |
12. | Kim WY, Suh HJ, Hong SB, Koh Y, Lim CM. Diaphragm dysfunction assessed by ultrasonography: influence on weaning from mechanical ventilation. Crit Care Med 2011; 39(12):2627–2630. |
13. | Boussuges A, Gole Y, Blanc P. Diaphragmatic motion studied by m-mode ultrasonography: methods, reproducibility, and normal values. Chest 2009; 135(2):391–400. |
14. | Yang KL, Tobin MJ. A prospective study of indexes predicting the outcome of trials of weaning from mechanical ventilation. N Engl J Med 1991; 324(21):1445–1450. |
15. | Similowski T, Yan S, Gauthier AP, Macklem PT, Bellemare F. Contractile properties of the human diaphragm during chronic hyperinflation. N Engl J Med 1991; 325(13):917–923. |
16. | Man WD, Soliman MG, Nikoletou D, Harris ML, Rafferty GF, Mustfa N et al. Non-volitional assessment of skeletal muscle strength in patients with chronic obstructive pulmonary disease. Thorax 2003; 58(8):665–669. |
17. | Donaldson AV, Maddocks M, Martolini D, Polkey MI, Man WD. Muscle function in COPD: a complex interplay. Int J Chron Obstruct Pulmon Dis 2012; 7:523–535. |
18. | Cohn D, Benditt JO, Eveloff S, McCool FD. Diaphragm thickening during inspiration. J Appl Physiol (1985) 1997; 83(1):291–296. |
[Figure 1], [Figure 2]
[Table 1], [Table 2]
This article has been cited by | 1 |
Accuracy of respiratory muscle assessments to predict weaning outcomes: a systematic review and comparative meta-analysis |
|
| Diego Poddighe, Marine Van Hollebeke, Yasir Qaiser Choudhary, Débora Ribeiro Campos, Michele R. Schaeffer, Jan Y. Verbakel, Greet Hermans, Rik Gosselink, Daniel Langer | | Critical Care. 2024; 28(1) | | [Pubmed] | [DOI] | | 2 |
The Role of Ultrasonography in the Process of Weaning from Mechanical Ventilation in Critically Ill Patients |
|
| Lou’i Al-Husinat, Basil Jouryyeh, Ahlam Rawashdeh, Chiara Robba, Pedro Leme Silva, Patricia Rieken Macedo Rocco, Denise Battaglini | | Diagnostics. 2024; 14(4): 398 | | [Pubmed] | [DOI] | | 3 |
The ratio of respiratory rate to diaphragm thickening fraction for predicting extubation success |
|
| Dararat Eksombatchai, Chalermwut Sukkratok, Yuda Sutherasan, Detajin Junhasavasdikul, Pongdhep Theerawit | | BMC Pulmonary Medicine. 2023; 23(1) | | [Pubmed] | [DOI] | | 4 |
Diaphragmatic Function Assessment Using Chest Ultrasonography as a Predictor for Weaning from Mechanical Ventilation |
|
| Tamer Sayed Abdel Mawla, Sherif Refaat Abdel Fattah, Afnan Mahmoud Abdel Halim, Radwa Ahmed Elhefeny | | Egyptian Journal of Critical Care Medicine. 2022; 9(1): 1 | | [Pubmed] | [DOI] | | 5 |
Ultrasound-assessed diaphragmatic dysfunction as a predictor of weaning outcome in mechanically ventilated patients with sepsis in intensive care unit |
|
| Mohamed Ahmed Saad, Sherif Wadie Nashed, Ahmed Nagah El-Shaer, Ashraf Elsayed Elagamy, Maha Sadek El derh | | Ain-Shams Journal of Anesthesiology. 2022; 14(1) | | [Pubmed] | [DOI] | | 6 |
World Experience Using of Soda Production Waste. |
|
| S.A. Krasilnikova, S.?. Blinov, P.?. Krasilnikov, P.?. Belkin | | Ecology and Industry of Russia. 2021; 25(12): 48 | | [Pubmed] | [DOI] | | 7 |
Diagnostic accuracy of diaphragm ultrasound to predict weaning outcome: A systematic review and meta-analysis |
|
| Aymeric Le Neindre, François Philippart, Marta Luperto, Johan Wormser, Johanna Morel-Sapene, Serge L. Aho, Silvia Mongodi, Francesco Mojoli, Belaid Bouhemad | | International Journal of Nursing Studies. 2021; 117: 103890 | | [Pubmed] | [DOI] | | 8 |
Diaphragm and Lung Ultrasound Indices in Prediction of Outcome of Weaning from Mechanical Ventilation in Pediatric Intensive Care Unit |
|
| Dalia A. Abdel Rahman,S. Saber,A. El-Maghraby | | The Indian Journal of Pediatrics. 2020; | | [Pubmed] | [DOI] | | 9 |
ICU outcomes can be predicted by noninvasive muscle evaluation: a meta-analysis |
|
| Clément Medrinal, Yann Combret, Roger Hilfiker, Guillaume Prieur, Nadine Aroichane, Francis-Edouard Gravier, Tristan Bonnevie, Olivier Contal, Bouchra Lamia | | European Respiratory Journal. 2020; 56(4): 1902482 | | [Pubmed] | [DOI] | | 10 |
Role of Bedside Ultrasonography in Assessment of Diaphragm Function as a Predictor of Success of Weaning in Mechanically Ventilated Patients |
|
| Mostafa Ibrahim Elshazly, Khaled Mahmoud kamel, Reem Ibrahim Elkorashy, Mohamed Said Ismail, Jumana Hesham Ismail, Hebatallah Hany Assal | | Tuberculosis and Respiratory Diseases. 2020; 83(4): 295 | | [Pubmed] | [DOI] | | 11 |
Diaphragm and weaning from mechanical ventilation: anticipation and outcome |
|
| RashaM Abdelhafeez,AhmedM Abumossalam,EmanO Arram,MohsenM Elshafey,MohammedE Abushehata | | Egyptian Journal of Bronchology. 2019; 13(4): 489 | | [Pubmed] | [DOI] | | 12 |
Diaphragm ultrasound as a predictor of successful extubation from mechanical ventilation |
|
| Taher Abd El Hamid El Naggar, Ibrahim A. Dwedar, Eman F. A. Abd-Allah | | Egyptian Journal of Bronchology. 2019; 13(2): 191 | | [Pubmed] | [DOI] | | 13 |
Prospective study: Diaphragmatic thickness as a predictor index for weaning from mechanical ventilation |
|
| Wafaa M. Abdelwahed,Mohamed S. Abd Elghafar,Yasser M. Amr,Salah El-din I. Alsherif,Mohamed A. Eltomey | | Journal of Critical Care. 2019; 52: 10 | | [Pubmed] | [DOI] | | 14 |
Diaphragmatic ultrasonography for predicting ventilator weaning |
|
| Caifeng Li,Xin Li,Hongqiu Han,Hailong Cui,Guolin Wang,Zhiqiang Wang | | Medicine. 2018; 97(22): e10968 | | [Pubmed] | [DOI] | | 15 |
Lung and diaphragm ultrasound as predictors of success in weaning from mechanical ventilation |
|
| Eva Tenza-Lozano,Ana Llamas-Alvarez,Enrique Jaimez-Navarro,Javier Fernández-Sánchez | | Critical Ultrasound Journal. 2018; 10(1) | | [Pubmed] | [DOI] | | 16 |
Diaphragm and Lung Ultrasound to Predict Weaning Outcome |
|
| Ana M. Llamas-Álvarez,Eva M. Tenza-Lozano,Jaime Latour-Pérez | | Chest. 2017; | | [Pubmed] | [DOI] | | 17 |
Diaphragmatic and lung ultrasound application as new predictive indices for the weaning process in ICU patients |
|
| Ahmed M. Osman,Reham M. Hashim | | The Egyptian Journal of Radiology and Nuclear Medicine. 2017; | | [Pubmed] | [DOI] | |
|
|
|
|